MERCK – 110274 Sterikon® Plus Bioindicator for Checks on Autoclaving

Specifications :

  • Appearance : opalescent and reddish violet
  • Number of viable spores per ampoule : 5.0E+05 – 1.0E+07
  • Growth (35 °C) : no growth
  • Growth (60 °C) : growth
  • Performance test (6 min, 121 °C) : Survival time 6 min/121 °C
  • Performance test (15 min, 121 °C) : Kill time 15 min/121 °C
  • D value (121 °C) : 1.5 – 2.0 min
  • Z value : 7.0 – 10.0 °C
  • Identity : Geobacillus stearo- thermophilus ATCC 7953
Kategori: Tag:

Deskripsi

MERCK – Sterikon® Plus Bioindicator for Checks on Autoclaving

Product Categories