HACH – TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System

Specifications :

Cable Length : 10 m (32.8 ft.) cable with marks for distance
Current Input : Approx. 60 mA
Dimensions : Case L: 450 mm (17.7 in)
: Meter (H X W X D): 110 x 230 x 40 mm (4.33 x 9.06 x 1.57 in.)
: Probe (Ø X L) 40 mm x 29 cm (1.57 x 11.42 in.)
Display : Alphanumerical display, 4 rows of 16 characters
Material : Sensor Body: stainless steel;
: Probe: stainless steel;
: Sensor Window: sapphire
Measurement Modes : Continuous measurement, integration time settings 0 – 1000 seconds
Measuring Principle : Combined multiple beam alternating light method with

: infrared diode system and beam focusing.;

: Turbidity: 2-channel 90° scattered light measurement in

: accordance with DIN/EN 27027/ISO 7027;

: additional measurement value verification through

: six-channel multiple-angle measurement;

: Solids: Modified absorption measurement;

: six-channel multiple-angle measurement

Measuring Range TSS : 0.001 to 400 g/L upper range depending on sample characteristics
Measuring Range Turbidity : 0.001 to 9999 FNU
Memory : Up to 290 measuring points
Operating Pressure : 0 to 10 bar (0 to 145 psi)
Operating Temperature Range : 0 to 60 °C

 

Kategori: Tag: ,

Deskripsi

HACH – TSS Portable Hand-held Turbidity, Suspended Solids, and Sludge Level System

Compact instrument for analytical determination by continuous measurement mode with integration time setting.
Adjustment of up to 5 measuring points, storage of measuring values with name, date and time. Shockproof.
Combined multiple beam alternating light method with infrared diode system and beam focusing; wavelength 860 nm. The measurement ranges for turbidity and suspended solids are: 0.001 – 9999 FNU and 0.001 – 400 g/L respectively.

  • Three Parameters in One Instrument
  • Multiple Calibration Curves for Convenience
  • Easy Sludge Blanket Levels
  • Air Bubble Compensation for Accuracy
  • Durable Materials for Long Life

Product Categories